Q: Is the autosampler temperature-controlled?



A: Yes! The autosampler is temperature-controlled. It can cool as low as 4 °C and heat up to 60 °C in order to cover a variety of samples and their stability or solubility requirements.




Q: What is the smallest volume I can inject if I don’t have much sample?



A: The OMNISEC autosampler can make accurate injections down to volumes of 5 µL. The autosampler can inject down to 1 µl, but without quite the same accuracy. Additionally, there are different injection options, such as a Zero Waste mode, which means you’re not wasting sample flushing out the sample loop.

Q: Can the software be used in batch processing mode, like the trend view was used in previous versions of OmniSEC software (v5 and before)?

A: The batch processing mode available in the previous versions of the OmniSEC software (v5 and below) is a feature that will soon be added to the OMNISEC v10 platform. 